Notice: Undefined variable: result in /www/wwwroot/careddipharma.com/system/storage/modification/catalog/controller/product/product.php on line 282Notice: Undefined variable: image in /www/wwwroot/careddipharma.com/system/storage/modification/catalog/controller/product/product.php on line 288 LC-MS/MS 6880 Triple Quadrupole Liquid Chromatography Mass Spectrometer
LC-MS/MS 6880 Triple Quadrupole Liquid Chromatography Mass Spectrometer

LC-MS/MS 6880 Triple Quadrupole Liquid Chromatography Mass Spectrometer

Product Code: LC-MS/MS 6880
Availability: In Stock
  • $0.00

  • Ex Tax: $0.00
LC-MS/MS 6880 Triple Quadrupole Liquid Chromatography Mass Spectrometer Product Parameter
Technical Parameters
Ultra high pressure liquid chromatograph
Pump system
dual-cylinder HPG (high-pressure mixing) dual-channel
Flow range
0.001 ~ 2.000mL/min (maximum 5 mL/min)
Flow accuracy
1% or 10uL/min
≤0.075% RSD or 0.005min SD
Maximum work pressure
0 ~ 18850psi (0 ~ 1300 bar)
Sample capacity
108 positions (2ml vial)
Sample temperature
4 ~ 40℃
Triple Quadrupole Mass Spectrometer
Ion source
1. Ion source ensures that number of ionization is maximized, also neutral particles and high-energy ion clusters that generate
noise are minimized.
2. Axial direction of spray atomizer is perpendicular to interface of orifice plate to improve spray effect and greatly reduce
entry of neutral particles into mass spectrometer, thereby eliminating noise.
3. By providing Electrospray Ionization (ESI) source and atmospheric pressure chemical ionization source (APCI), substances of
different polarities can be analyzed.
Ion guide
High frequency (RF) double ion funnel is optically optimized to transport a wide range of ions while performing efficient ion
focusing.
Quadrupole and front and rear filters 1 and 2 are used to optimize ion transmission and spectral level resolution.
Collision unit
Collision unit has optimized mode of high-pressure 180-degree bending, and at same time eliminates interference.
Detector
Electron accelerator with acceleration after 5kV.
Vacuum system
Two single-stage 40L/h primary pumps
Three-stage 25/300/400L/s turbomolecular pumps
System performance
1. MRM sensitivity, signal-to-noise ratio (S/N), ESI positive ion
1pg reserpine, scan m/z 609 to 195 for quantification.
S/N: 1000000:1
2. MRM sensitivity, signal-to-noise ratio (S/N), ESI negative ion
1pg chloramphenicol, scan m/z321 to 152 for quantification.
S/N: 500000:1
3. Mass range: m/z 10 ~ 1250
4. Polarity switching: 25ms
5. Mass resolution: 0.7 Da to 4.0 Da (user adjustable)
6. Quality stability: 0.1Da/24h
7. Dynamic range: >105
8. Scan mode: full scan, product ion scan, precursor ion scan, neutral loss scan, selective ion scan (SRM/MRM)
9. Maximum scanning rate: 20000 Da/S
10. Maximum MRM collection rate: 600 MRMs/S
11. Minimum MRM residence time: 1.0 ms

INTRODUCTION:

LC-MS/MS 6880 Triple Quadrupole Liquid Chromatography Mass Spectrometer simplifies the complex analysis process through an intelligent workflow, and has high sensitivity and reliable quantitative functions. Each individual module can be operated
intelligently. The system can provide lower quantitative detection limit and selectivity of synchronous data processing, thereby
improving the maximum laboratory efficiency. It provides the best mass spectrometry solutions and intelligent instrument
configuration for biopharmaceuticals, universities, research institutes and other institutions.

PERFORMANCE CHARACTERISTICS

* ESI/APCI free switch, optional installation
* High-efficiency ceramic heater heat setting technology
* Automatic emptying system to reduce the interference to ion source pollution
* Instrument man-machine dialogue access mode, more intelligent
* The best overall solution
* Reliable quantitative data
* Comprehensive real cases of the project
* Fast report mode
* Software specialization and unique interface protocol provide reliability
* Quality Express TM, Data Express TM,
* Unique professional extension interface of the software
* Maintain high sensitivity
* Simplify the work process
* Improved work efficiency
* Ensure the stability of the instrument
* Enhance the ionization efficiency, reduce the degradation of thermally unstable compounds, and increase sensitivity.

APPLICATION DIRECTION

High-performance instruments can provide the most ideal solution, which can be widely used in food safety, drug analysis,
environmental science, toxicological judgment, and compound structure qualitativeness.